Crossref journal-article
AIP Publishing
Review of Scientific Instruments (317)
Abstract

A modular system of techniques and software has been developed for the calibration and correction of intensity linearity, uniformity of response, spatial distortion, and image plate decay. With calibration the Molecular DynamicsTM Imaging Plate scanner system has been shown to give comparable results to the MarResearchTM scanner. The ESRF x-ray image intensifier/charge-coupled device detectors inherently cause large spatial and uniformity of response distortions, and successful data analysis depends on calibration and correction. Results of synchrotron radiation experiments are presented.

Bibliography

Hammersley, A. P., Svensson, S. O., Thompson, A., Graafsma, H., Kvick, Å., & Moy, J. P. (1995). Calibration and correction of distortions in two-dimensional detector systemsa). Review of Scientific Instruments, 66(3), 2729–2733.

Authors 6
  1. A. P. Hammersley (first)
  2. S. O. Svensson (additional)
  3. A. Thompson (additional)
  4. H. Graafsma (additional)
  5. Å. Kvick (additional)
  6. J. P. Moy (additional)
References 15 Referenced 138
  1. 10.1098/rspa.1990.0030 / Proc. R. Soc. London Ser. A (1990)
  2. 10.1107/S0021889892004035 / J. Appl. Crystallogr. (1992)
  3. {'key': '2024021017325788600_r3'}
  4. 10.1016/0168-9002(94)90815-X / Nucl. Instrum. Methods A (1994)
  5. 10.1117/12.142034 / Proc. SPIE (1993)
  6. 10.1016/0168-9002(90)90829-U / Nucl. Instrum. Methods A (1990)
  7. 10.1098/rspa.1989.0102 / Proc. R. Soc. London Ser. A (1989)
  8. 10.1016/0168-9002(94)90720-X / Nucl. Instrum. Methods A (1994)
  9. {'key': '2024021017325788600_r9', 'volume': '29', 'year': '1993', 'journal-title': 'Newsletter'} / Newsletter (1993)
  10. 10.1117/12.59393 / Proc. SPIE (1992)
  11. 10.1016/0168-9002(91)91029-U / Nucl. Instrum. Methods A (1991)
  12. {'key': '2024021017325788600_r12', 'first-page': '137', 'volume': '47', 'year': '1991', 'journal-title': 'Acta Crystallogr. B'} / Acta Crystallogr. B (1991)
  13. 10.1016/0014-5793(94)01060-9 / FEBS Lett. (1994)
  14. 10.1107/S0108767390000277 / Acta Crystallogr. A (1990)
  15. 10.1107/S0567740876006924 / Acta Crystallogr. B (1976)
Dates
Type When
Created 23 years, 1 month ago (July 26, 2002, 8:29 a.m.)
Deposited 1 year, 6 months ago (Feb. 10, 2024, 1:09 p.m.)
Indexed 2 days, 13 hours ago (Aug. 26, 2025, 2:38 a.m.)
Issued 30 years, 5 months ago (March 1, 1995)
Published 30 years, 5 months ago (March 1, 1995)
Published Print 30 years, 5 months ago (March 1, 1995)
Funders 0

None

@article{Hammersley_1995, title={Calibration and correction of distortions in two-dimensional detector systemsa)}, volume={66}, ISSN={1089-7623}, url={http://dx.doi.org/10.1063/1.1145618}, DOI={10.1063/1.1145618}, number={3}, journal={Review of Scientific Instruments}, publisher={AIP Publishing}, author={Hammersley, A. P. and Svensson, S. O. and Thompson, A. and Graafsma, H. and Kvick, Å. and Moy, J. P.}, year={1995}, month=mar, pages={2729–2733} }